<4. PARAMETER SETTING>
4-2
IM 12E01A06-02E
4.2 Operation Level
This is the level where, basically, operations regarding routine inspections and maintenance, such as
calibration and manual cleaning, can be done. Also, the desired display item on the message display
can be selected at this operation level.
F4-2E.ai
[YES]
[MODE]
[NO]
[YES]
[YES]
[NO]
[YES]
[YES]
[YES]
[YES]
[YES]
[NO]
[NO]
HOLD
CALIB
Calibration mode
DISP
Message display selection mode
HOLD
SETP
Alarm setting mode
HOLD
Hold selection mode
Measurement
mode
*1: Skipped if alarm setting function is set to “0: Disabled” in Code 51 at service level, or
unless either of functions of contact output S1 or S2 is set to “1: Alarm” in Code 40 or 41
at service level.
*2: Skipped if hold function is set to “*H.OFF: Disabled” in SET HOLD mode at setting level.
*3: If [YES] key is pressed, analog output is held in measurement mode.
*1
*2
*3
Table 4.1
Setting Parameters at Operation Level
Mode
Setting Parameter
Display
Setting Range
Default
CAL
Calibration
CALIB
Using turbidity standards,
or using water sample
STD.CAL or
SMP.CAL
Start
ZERO or SPAN
or ZRSPN
Turbidity value
VALUE
0.001 to 2000[NTU]
*1
Self-adjustment in process
ZERO or SPAN
End
CAL.END
CAL START
Not available
WASH START
Not available
DISPLAY
Message display selection
DISP
Analog output 1 current
XX.XmA1
Output 1 current
Analog output 2 current
XX.XmA2
Switchable output range
state
RNG1/2-A/B/C
*2
(No pointer at
Alarm setting
* SETP
*3
mode display) High alarm setpoint
* S.TB-H
–010.0 to 2200.0[NTU] *1
2200.0
(NTU)
Low alarm setpoint
* S.TB-L
–010.0 to 2200.0[NTU] *1
-010.0
(NTU)
HOLD
Hold selection
HOLD
*4
Hold selection
(Disabled)
*1:
Although setting range is up to 2000 NTU, measuring range is up to 100 NTU.
*2:
Skipped if analog output range switching selection {*RNGPR} is set to “0: Fixed” in CODE 30 at service level.
*3:
Skipped if alarm setting function is set to “0: Disabled” in Code 51 at service level, or if either function of contact output S1 or S2
is not set to “1: Alarm” in Code 40 or 41 at service level.
*4:
Skipped if hold function is set to “*H.OFF: Disabled” in SET HOLD mode at setting level.