<6. TROUBLESHOOTING>
6-1
IM 12E01A06-02E
6. TROUBLESHOOTING
Table 6.1
Error Code List (1/3)
No.
Error
Occurrence
Problem
Return
E102
EEPROM write
failure
All modes
Data is written in EEPROM and verified,
and write failure occurs 3 consecutive
times.
When power is
turned off and then on
again, and problem is
eliminated. (Without
[YES]/[NO] key press.)
E103
RAM failure
When power is
turned on.
RAM area failure (3 consecutive times).
E104
AD converter
failure
All modes
AD converter failure (3 consecutive
times).
E201
Input voltage
failure
All modes
Either IN1 or IN2 input voltage is outside
the range of -0.15 to 1.2 V (fixed),
including abnormal value 0x7FFF or
8000, for 5 consecutive seconds.
When problem is
eliminated for 5
consecutive seconds.
When [YES]/[NO] key
is pressed, error code
display disappears, at
the same time during
failure, error is once
removed.
E202
Detection
element failure
All modes
Either IN1 or IN2 input voltage is less
than specified voltage for 5 consecutive
seconds.
E203
Lamp life
expired
Zero calibration
coefficient A
rewriting timing.
(excluding when
E301 or E311
occurs.)*
1
During zero calibration self-adjustment,
input voltage IN1 is less than specified
voltage for 5 consecutive seconds.
(Not detected when E301 is occurring.)
Rewriting of zero calibration coefficient A
is performed.
E204 Lamp intensity
failure
E205
Calibration
failure
E301 to E307,
E311 to E317,
E321 during PC
communication
occurrence
timing
Any of E301 to E307, E321 during PC
communication is occurring.
When problem is
eliminated. When
[YES]/[NO] key is
pressed, error code
display disappears, at
the same time during
failure, error is once
removed.