Transmitter
General Instruction Manual
Wiring
IM 01U10B00-00EN-R, 3rd edition, 2018-07-09
59 / 90
Active status
output
P/Sout
Since this is a transistor contact, maximum allowed current as well as polarity and level of
output voltage must be observed during wiring.
Value
Load resistance
> 1 kΩ
Internal power supply
24 V
DC
±20 %
P/Sout+
P/Sout-
24 V
0 V
ROTAMASS
1
Fig. 46:
Active status output connection
P/Sout
①
External device with load resistance
Active status
output
P/Sout
with internal
pull-up resistor
Value
Internal pull-up resistor
2.2 kΩ
Internal power supply
24 V
DC
±20 %
P/Sout+
P/Sout-
24 V
0 V
ROTAMASS
1
Fig. 47:
Active status output
P/Sout
with internal pull-up resistor
①
External device