<6. ELECTRONIC TRANSMITTER (-E)>
6-6
IM 01R01B02-00E-E 12th edition October 01, 2014 -00
All Rights Reserved. Copyright © 2003, Rota Yokogawa
With functions F12 and F13, the measuring unit for displayed value is selected.
F12 selects volume and mass units, while F13 sets the corresponding time unit.
When selecting the indication function “totalizer” the set time unit is not taken into account and only the
selected mass or volume unit is effective. When choosing the “%” indication F12 and F13 have no effect.
The selection of the measuring unit is performed as follows:
Description
Selection
Key
Indication
Change to setting mode
Enter
Display mode
F1-
Setting
Masse/Volume unit
unit set
Euro US
Enter
→
Enter
F11
F12
F12 -1
Selection unit
or
or
or
or
or
or
m3
m3
l
acf
Nm3 Nm3
Nl
scf
t
ton
kg
kg
scf
lb
gal
usg
Enter
→
Enter
2 x
→
Enter
3 x
→
Enter
4 x
→
Enter
5 x
→
Enter
6 x
→
Enter
7 x
→
Enter
F12
F12 -2
F12
F12 -3
F12
F12 -4
F12
F12 -5
F12
F12 -6
F12
F12 -7
F12
F12 -8
F12
Setting
Time unit
→
Enter
F13
F13 -1
Selection time unit
or
or
or
h
h
min.
min
s
s
day
day
Enter
→
Enter
2 x
→
Enter
3 x
→
Enter
F13
F13 -2
F13
F13 -3
F13
F13 -4
F13
Back to display mode
↑
↑
F1-
Display mode
Note:
If you press “
↑
" instead of “Enter”, you can return from the selected point to the previous menu
without activating the displayed parameter.
After changing the measuring unit the corresponding measuring unit label should be fixed on the right side
next to the display. A sheet with stickers is included.
Attention:
When switching the mass/volume unit the totalizer is reset to zero.
When changing the time unit the totalizer value remains unchanged.