<6. ELECTRONIC TRANSMITTER (-E)>
6-12
IM 01R01B02-00E-E 12th edition October 01, 2014 -00
All Rights Reserved. Copyright © 2003, Rota Yokogawa
6.2.8 Error messages (F4-)
I
f the 8 bars beneath the digits start flashing, an error has occurred in the measuring transmitter/current
output. Since the pointer indication is independent from the electric measuring transmitter, it may show the
correct measuring value even if the transmitter is defective. Function F41 allows checking of assigned error
codes.
Error codes are called onto the display as follows:
Description
Selection
Key
Indication
Change to setting mode
Enter
Display mode
F1-
Setting mode
3 x
→
Enter
Enter
Enter
F4-
F41
F41 Enn
F41
Back to display mode
↑
↑
F4-
Display mode
List of error messages :
Code
Meaning
Remedy
01
02
03
04
05
06
07
08
RAM-error
ADC-error
Internal EEPROM faulty
Calibration-EEPROM faulty
Wrong totalizer value in EEPROM
Overflow (flow rate too high)
Internal EEPROM faulty
Float blocking indication realized,
supervision time gone off
Indication unit needs service
Indication unit needs service
Indication unit needs service
If EEPROM is missing insert, -otherwise order new
EEPROM
Reset totalizer
Reduce flow rate
Indicator unit needs service
Deactivate float-blocking-indication or run Autozero
function
In case of error the appropriate remedy has to be taken.