6.2.7.4 Calculation of pulse rate
Concerning the final flow-value (Qmax ), which was declared in the customers order, the pulse rate is
calculated in factory and is written onto the sheet ´Data of Pulse Output (Option /CP)´, which is included. This
value can be transferred to a blank label of the also included sheet with unit-stickers and then fixed on the
scale. After changing the flow rate unit with F12 the pulse rate must be recalculated.
Calculation of pulse rate :
· Read the value of Qmax from the scale or recalculate it.
· Search for the concerning range in the first row of the table below.
· Read the concerning pulse rates in the second and third row.
· The measuring unit is equal to the flow
Maximum flow Qmax
without unit
Pulse rate for F34-2
without unit
Pulse rate for F34-3
without unit
Qmax
≤
1
1 < Qmax
≤
10
10 < Qmax
≤
100
100 < Qmax
≤
1000
1000 < Qmax
≤
10000
10000 < Qmax
≤
100000
0.0001
0.001
0.01
0.1
1
10
0.001
0.01
0.1
1
10
100
e.g..: Final value (Qmax ) = 400 m3/h
→
Pulse rate 0.1 m
3
for F34-2 and 1 m
3
for F34-3
•
The factory default is F34 -2 (solution : 1 last digit).
•
After Master Reset F34 -1 (Pulse output not active) is selected.
•
After changing the flow rate unit with F12 the pulse-register is deleted and the pulse rate is automatically
redefined according to the new unit.
•
After totalizer Reset with F14 the pulse output does not change.
•
When changing the indication with F11 -3 to ´%´, the totalizer stops and the pulse output is switched off.
•
After switching power on one pulse is generated at the output.
•
For 2- respectively 3-wire-units the function F34 is not supported.