SO/SO?
Sets whether to start or stop data storage in
auto store mode/queries the current setting.
Syntax
SO m <terminator>
“m” represents whether data storage is to
be started or stopped.
0: Stops (aborts) data storage.
1: Starts data storage.
Query
SO? <terminator>
Response Example
SO0
Description
• Cannot execute in manual store mode.
An error 14 occurs.
• Cannot change in recall mode. An
error 14 occurs.
ST
Stores data in manual store mode.
Syntax
ST <terminator>
Description
• Cannot execute in auto store mode.
An error 14 occurs.
• Cannot execute in recall mode. An
error 14 occurs.
SY/SY?
Sets DMM calibration mode and normal
measurement mode and queries the current
setting.
Syntax
SYm<terminator>
“m” represents the system mode.
0: Normal measurement mode
1: DMM calibration mode
Query
SY? <terminator>
Response Example
SY0
Description
• DMM calibration mode cannot be set if
the DMM function is OFF. An error 16
occurs.
• DMM calibration mode cannot be set
when data is on hold. An error 15
occurs.
• DMM calibration mode cannot be set
when in recall mode. An error 14
occurs.
VO/VO?
Turns 24 VDC output ON and OFF and queries
the current setting.
Syntax
VO m <terminator>
“m” represents ON/OFF of 24 VDC
output.
0: 24 VDC output is OFF
1: 24 VDC output is ON
Query
VO? <terminator>
Response Example
VDO