Appendix
App-5
IM 767361-01E
Syntax
DB m, n <terminator>
“m” represents the Number of the first
memory location to be deleted.
m=1 to 9999
“n” represents the number of the last
memory location to be deleted.
n=1 to 9999 (n≥m)
Query
DB? <terminator>
Response Example
DB1, 1
Description
• If the end memory number is smaller
than the start memory number, error 12
occurs.
• Cannot be executed during auto store.
An error 14 occurs.
DF/DF?
Makes the current/voltage measurement setting/
queries the current setting.
Syntax
DF m <terminator>
“m” represents the measurement item.
1: DC voltage (5V range)
2: DC current (20mA range)
Query
DF? <terminator>
Response Example
DF1
Description
• Cannot change during data hold. An
error 15 occurs.
• Cannot change with DMM function
OFF. An error 16 occurs.
• Cannot change in recall mode. An
error 14 occurs.
DIS/DIS?
Sets display mode and queries the current
setting.
Syntax
DIS m <terminator>
“m” represents the display mode.
0: Normal measurement display
1: % display
2: %ERROR display (possible only when
the DMM function is ON)
Query
DIS? <terminator>
Response Example
DIS0
Description
Cannot be changed while displaying the
relative value. An error 15 occurs.
DL/DL?
Sets the communications output data terminator
and queries the current setting.
Syntax
DLm<terminator>
“m” represents the terminator type.
GP-IB
RS-232
0: CR+LF+EOI CR+LF
1: LF
LF
2: EOI
CR
Query
DL?<terminator>
Response Example
DL0
DMM/DMM?
Turns the DMM function ON and OFF and
queries the current setting.
Syntax
DMM m <terminator>
“m” represents ON/OFF for the DMM
function.
0: DMM function is OFF
1: DMM function is ON.
Query
DMM? <terminator>
Response Example
DMM1
Description
• Cannot change during data hold. An
error 15 occurs.
• Cannot change in recall mode. An
error 14 occurs.
Appendix