Overview of the GP-IB Interface
53
IM 767361-01E
Overview of the GP-IB Interface
Address Setting Range
0–30
The factory setting is 1.
Talk-Only Function
This function only allows the instrument to send measured or computed data
to other devices. Even if the instrument is not assigned as a talk-only device,
it can still send data. In talk-only mode, the instrument cannot be controlled by
the controller.
Setting the GP-IB Interface
1.
Press
MENU
.
Keep pressing or until
is displayed in the lower part of the screen,
then press ENTER/LOCAL.
Proceed to step 2 when using the battery pack or DC power supply.
Proceed to step 3 when using the AC power supply.
2.
Use the and to set the GP-IB function (
) and press
ENTER/LOCAL
.
: Activates the function.
: Deactivates the function.
3.
Press or to select
(addressable mode) or
(talk-only mode),
then confirm the selection by pressing
ENTER/LOCAL
.
4.
If addressable mode is selected, keep pressing until the desired digit starts to
blink, then set the desired value by pressing or to set the address (
),
then press
ENTER/LOCAL
.
5.
Press
MENU
to return to the normal measurement screen.