<13. MEASUREMENT PRINCIPLE>
13-2
IM 11P03A03-01E
6th Edition : Jan. 26, 2022-00
For example, if carbon dioxide which has a smaller susceptibility than nitrogen is passed through
the cell, the indication deflects to a negative value. If the cell is tilted (angled) to compensate the
indication as shown in Figure 13.3, a counterforce acts to supply more auxiliary gas into path A
due to the higher density of carbon dioxide, thus changing the flow ratio of the auxiliary gas. By
changing the cell angle so that the change in the auxiliary gas flow ratio due to gas susceptibility
cancels out the flow ratio change due to the density difference, the interference error can be
compensated.
The model MG8 paramagnetic oxygen analyzer is shipped after adjusting the cell inclination in
the final adjustment stage using the magnetic characteristics and density of the measuring gas of
the user. In this case, the inclination is stored using a built-in level (containing a bubble in a glass
tube).
When this analyzer is installed at the user’s site, turning the rotary knob so that the bubble of this
level returns to the center reproduces the adjustment when shipped.
Magnetic field
Thermistor
Stream B
Stream A
Gas outlet
Auxiliary gas inlet
Measurement gas inlet
F16.2E.ai
Figure 13.2
Top of Cell
Stream B
Stream A
Auxiliary gas
Inclination
F16.3E.ai
Thermistor
Figure 13.3
Cross Section of Cell
l
How to adjust the inclination
If the attitude must be adjusted differently, e.g., if the composition of the measuring gas changes,
follow the procedure described in Chapter 8, “ PREPARATIONS” and the angle adjustment
procedure for interference gas compensation.
The model MG8 analyzer allows the angle to be adjusted from outside of the explosion-protected
enclosure in order to preserve the explosion-protected condition. (However, since the analyzer
scarcely requires inclination adjustment for an MS code with a span of 5 % or more, adjustment
is done by opening the explosion-protected door, if necessary.)
Содержание MG8G
Страница 1: ...User s Manual Model MG8G Paramagnetic Oxygen Analyzer IM 11P03A03 01E IM 11P03A03 01E 6th Edition...
Страница 5: ...Blank Page...
Страница 9: ...Blank Page...
Страница 13: ...Blank Page...
Страница 17: ...Blank Page...
Страница 21: ...Blank Page...
Страница 27: ...Blank Page...
Страница 61: ......
Страница 63: ...Blank Page...