<8. PARAMETER SETTINGS>
8-4
IM 11P03A03-01E
6th Edition : Jan. 26, 2022-00
8.6
Setting the Range Switching Answerback
Contact Output
The range switching answerback function enables the currently selected range to be output
by the contact output (terminals 10, 11 and 12). Since either the range switching answerback
function or the high/low limit alarm function can be assigned to this contact, the range switching
answerback function cannot be used when the high/low limit alarm function is enabled.
(1) Switch to MAINT mode and enter the password.
(2) Change the Function Number to “22.”
(3) Set to “0” to disable the high/low limit alarm function using the SHIFT and INCR keys. Press
the EXEC key to accept. Now the range switching answerback function is enabled.
The range answerback contact states and ranges are shown in Figure 8.2.
Table 8.2
Range Answerback Contact States and Ranges
Terminal Number
Range Selected
10-12 (R3)
11-12 (R2)
Open
Open
Range 1
Open
Closed
Range 2
Closed
Open
Range 3
CAUTION
When the range answerback function is enabled, the high/low limit alarm function is disabled.