<6. OPERATIONS>
6-5
IM 11P03A03-01E
6th Edition : Jan. 26, 2022-00
6.3.6
Selecting Data
This section explains how to select and set one of parameters. As an example, selection of the
range is shown below.
Key Operation
FUNC
DATA
Procedure
Switch to MAINT mode.
Enter the password.
ENABL
INHIBIT
SPAN
ZERO
▲
FUNC
▲
DECIMAL
POINT
SHIFT
▲
INCR
EXEC
MEAS/MAINT
Press FUNC keys to show “11” on FUNC
display.
DATA display shows “1”, the currently
selected range.
ENABL
INHIBIT
SPAN
ZERO
▲
FUNC
▲
DECIMAL
POINT
SHIFT
▲
INCR
EXEC
MEAS/MAINT
Press SHIFT key.
“1” blinks, indicating that the digit can be
changed.
ENABL
INHIBIT
SPAN
ZERO
▲
FUNC
▲
DECIMAL
POINT
SHIFT
▲
INCR
EXEC
MEAS/MAINT
Press INCR key to change the digit to the
desired range number. The digit cycles
through only the available values (in this
case 1 through 3).
ENABL
INHIBIT
SPAN
ZERO
▲
FUNC
▲
DECIMAL
POINT
SHIFT
▲
INCR
EXEC
MEAS/MAINT
Press EXEC key.
The digit stops blinking, indicating that the
change has been accepted.