<5. PREPARATIONS>
5-1
IM 11P03A03-01E
6th Edition : Jan. 26, 2022-00
5. PREPARATIONS
5.1
Adjustment of the Level in the Sensor Unit
After the installation is complete, adjust the angle of the sensor unit.
(1) Loosen the four screws (
, these screws cannot be removed), and remove the constant
temperature chamber cover (
).
(2) Remove the six screws (
, 3 each on top and bottom), and remove the heater assembly
cover (
).
(3) Loosen the four bolts (
) with the supplied wrench.
(4) Place the supplied mirror (
) as shown in Figure 5.1 (b
). Turn the adjustment screw (
)
with a flat head screwdriver or other suitable tool until the air bubble in the level (
) comes
in between the two red lines. Turning the adjustment screw counterclockwise brings the
bubble position toward the front of the instrument.
(5) Tighten the four bolts (
) with the wrench. Make sure that the bubble is in position.
(6) Replace the heater assembly cover (
), and tighten the cover screws securely. Replace
the constant temperature chamber cover (
), and tighten the cover screws securely. When
replacing the covers, be sure that cables are not caught in the covers. Make sure that the
covers are installed securely. Otherwise, the temperature in the constant temperature
chamber may become unstable, causing measurement errors.
45
°
Mirror (supplied)
Level
Bolts, 4
×
M6
F8.2E.ai
F8.1E.ai
Constant Temperature
Chamber Cover
Bolts, 4×M6
Screws, 4×M4
Adjustment Screw
Level
Heater Assembly
Cover
Screws, 6×M4
Figure 5.1
(a) Constant Temperature Chamber: Front View
(b) Constant Temperature Chamber: Side View
Содержание MG8G
Страница 1: ...User s Manual Model MG8G Paramagnetic Oxygen Analyzer IM 11P03A03 01E IM 11P03A03 01E 6th Edition...
Страница 5: ...Blank Page...
Страница 9: ...Blank Page...
Страница 13: ...Blank Page...
Страница 17: ...Blank Page...
Страница 21: ...Blank Page...
Страница 27: ...Blank Page...
Страница 61: ......
Страница 63: ...Blank Page...