<4. Piping>
4-2
IM 11M12A01-02E 16th Edition : Mar. 19, 2020-00
• It is recommended to use ZH21B dust protector to protect the probe output from dust
agitation (i.e., to prevent combustible materials from entering the probe cell) where humidity
measurements are made under dusty or combustible, such as paper dust, environment.
•
If a high temperature detector is used and no piping can be installed for the reference gas,
place piping in the exhaust hole for the sample gas on the high temperature probe adapter
so that the sample gas is carried away from the vicinity of the detector (see Section 4.1.4,
Figure 4.6).
•
If a high temperature detector is used and the sample gas pressure is negative, connect
an ejector assembly to the sample gas exhaust hole of the high temperature probe adapter
(see Section 4.1.4, Figure 4.3).
•
If a high temperature detector is used and the pressure of the sample gas is 0.49 kPa or
higher, it is recommended that a needle valve (throttle) be used in the sample gas exhaust
of the high temperature probe adapter (see Section 4.1.4, Figure 4.4).
CAUTION
This is for lowering the sample gas temperature below 700°C. If the gas temperature is high and
the pressure is also significantly high, the sample gas temperature may not fall to below 700°C
before it reaches the detector.
On the other hand, if the sample gas temperature is lowered too much, condensation may be
produced in the High Temperature Probe Adapter. During wintertime, it is recommended that the
High Temperature Probe Adapter be protected with an insulating material to avoid condensation
forming (see Section 4.1.4, Figure 4.5).
For the usage of the High Temperature Probe Adapter, refer to Section 3.2.1.
4.1.1
Parts Required for Piping in System 1
Check that the parts listed in Table 4.1 are ready.
Table 4.1
Detector
Piping location
Parts Name
Note
General use
detector
Calibration gas inlet Stop valve
Recommended by YOKOGAWA
(L9852CB or G7016XH)
Nipple *
Rc 1/4 or 1/4 NPT
General parts
Joint for tube connection Rc 1/4 (1/4 NPT)
for a Ø6 x Ø4 mm soft tube
General parts
Reference gas inlet (sealed up)
(when piping is required, refer to section 4.1.3)
High temperature
detector
(0.15 m)
Calibration gas inlet Stop valve
Recommended by YOKOGAWA
(L9852CB or G7016XH)
Nipple *
Rc 1/4 or 1/4 NPT
General parts
Joint for tube connection Rc 1/4 (1/4 NPT) for a Ø6 x Ø4 mm soft tube
Reference gas inlet (sealed up)
(when piping is required, refer to section 4.1.3)
Sample gas outlet
Ejector assembly *
Recommended by YOKOGAWA
(E7046EC or E7046EN)
T-shaped joint of the
same diameter *
R1/4 or 1/4 NPT
General parts
Needle valve *
Rc1/4 or 1/4 NPT
General parts
Nipple of other diameter
*
R 1/2 to R 1/4 or
R 1/2 to 1/4 NPT
General parts
Note: Parts with marking * are used when required.
General parts can be found on the local market.
Содержание EXAxt ZR22G
Страница 15: ...Blank Page...
Страница 51: ...Blank Page...
Страница 77: ...Blank Page...
Страница 137: ...Blank Page...
Страница 171: ...Blank Page...
Страница 195: ...Blank Page...
Страница 197: ......
Страница 199: ......
Страница 201: ......
Страница 203: ......
Страница 205: ......
Страница 207: ......
Страница 211: ...Blank Page...