<2. Specifications>
2-14
IM 11M12A01-02E 16th Edition : Mar. 19, 2020-00
Semi-auto Calibration; Requires the ZR40H Automatic Calibration Unit. Input calibration
direction on the touchscreen or contact, then it calibrates
automatically afterwards.
Manual Calibration; Calibration with opening/closing the valve of calibration gas in
operation interactively with an LCD touchscreen.
Blow back Function:
Output through the contact in the set period and time. Auto/Semi_Auto selectable.
Maintenance Functions:
Can operate updated data settings in daily operation and checking. Display data settings,
calibration data settings, blow back data settings, current output loop check, input/output
contact check.
Setup Functions:
Initial settings suit for the plant conditions when installing the converter.
Equipment settings, current output data settings, alarm data settings, contact data
settings, other settings.
Self-diagnosis:
This function diagnoses conditions of the converter or the detector and indicates when
any abnormal condition occurs.
Password Functions:
Enter your password to operate the analyzer excepting data display.
Individual passwords can be set for maintenance and setup.
Display and setting content:
(inclused Humidity Analyzer)
Measuring related items: Oxygen concentration (vol%O
2
), output current value (mA), air ratio,
moisture quantity (in hot gases) (vol%H
2
O), mixture ratio (kg/kg),
relative humidity (%RH) and dew point (°C)
Display Items:
Cell temperature (°C), thermocouple reference junction temperature (°C),
maximum/minimum/average oxygen concentration (vol%O
2
), maxi-mum/
minimum/average moisture quantity (vol% H
2
O), maximum/minimum/
average mixture ratio (kg/kg), cell e.m.f. (mV), cell internal resistance (Ω),
cell condition (in four grades), heater on-time rate (%), calibration record
(ten times), time (year/month/day, hour/minute)
Calibration Setting Items: Span gas concentration (vol%O
2
), zero gas concentration (vol%O
2
),
calibration mode (auto, semi-auto, manual), calibration type and
method (zero-span calibration, zero calibration only, span calibration
only), stabilization time (min. · sec.), calibration time (min. · sec.),
calibration interval (day/hour), starting time (year/month/day, hour/
minute)
Equipment Related Items: Measuring gas selection
Output Related Items: Analog output/output mode selection, output conditions when warming-
up/maintenance/calibrating (during blow back)/abnormal, oxygen
concentration at 4 mA/20 mA (vol% O
2
), moisture quantity at 4 mA/20
mA (vol% H
2
O), mixture ratio at 4 mA/20 mA (kg/ kg), time constant.
Alarm Related Items: Oxygen concentration high alarm/high-high alarm limit values (vol%O
2
),
oxygen concentration low alarm/low-low alarm limit values (vol%O
2
),
moisture quantity high alarm/high-high alarm limit values (vol% H
2
O),
moisture quantity lowalarm/ low-low alarm limit values (vol% H
2
O),
mixture ratio high alarm/high-high alarm limit value (kg/ kg), mixture ratio
low alarm/low-low alarm limit values (kg/ kg), oxygen concentration alarm
hysteresis (vol%O
2
), oxygen concentration/moisture quantity/mixture
ratio alarm detection, alarm delay (seconds)
Содержание EXAxt ZR22G
Страница 15: ...Blank Page...
Страница 51: ...Blank Page...
Страница 77: ...Blank Page...
Страница 137: ...Blank Page...
Страница 171: ...Blank Page...
Страница 195: ...Blank Page...
Страница 197: ......
Страница 199: ......
Страница 201: ......
Страница 203: ......
Страница 205: ......
Страница 207: ......
Страница 211: ...Blank Page...