<7. Startup>
7-16
IM 11M12A01-02E
Enter
Zero calibration
Close the zero gas valve.
Span calibration
End
Manual calibration
F7.26E.ai
Figure 7.36
Zero Calibration Complete Display
(10) Select “End” in the display as shown in Figure 7.36. An oxygen concentration trend graph
(with the oxygen concentration being measured) appears and HOLD TIME then flashes.
This time is referred to as the output-stabilization time. If the HOLD TIME has been set
with the output hold setting, the analog output remains held (refer to Section 8.2, Output
Hold Setting, later in this manual). Manual calibration is completed when the preset hold
(output stabilization) time elapses. This hold (output stabilization) time is set to 10 minutes
at the factory before shipment. If you press the [Enter] or [Reject] key within the hold (output
stabilization) time, manual calibration is then completed.