<8. Detailed Data Setting>
8-21
IM 11M12A01-02E 16th Edition : Mar. 19, 2020-00
Enter
Temperature input mode:
mA-input
Input temperature value:
3
0
0
ºC
Pressure: 1 0 1. 3 3 kPa
Exhaust gas setup
F8-20E.ai
Enter
Temperature input mode:
mA-input
Temperature at 4mA:
0
ºC
Temperature at 20mA:
1 0 0 0
ºC
Alarm value of
temperature: 1 0 0 0
ºC
Pressure: 1 0 1 . 3 3 kPa
Exhaust gas setup
F8-21E.ai
Figure 8.19
Exhaust Gas Setup
Figure 8.20
Exhaust Gas Setup Display
Display (Preset Entry)
(Entering mA-input)
l
Setting Measurement Gas Pressure
To set the measurement gas pressure, follow these steps:
(1) Call up the Exhaust gas setup display and choose Pressure.
(2) Enter the measurement gas pressure (absolute pressure) from the numeric data entry display.
l
Default Values
When the analyzer is delivered, or if data are initialized, measurement gas temperature and
pressure set ranges and their default settings are as shown in Table 8.11.
Table 8.11
Measurement Gas Temperature and Pressure Set Ranges and Default Settings
Set Item
Set range
Default setting
Temperature input selection
Preset value
Measurement gas temperature
0 to 3000°C
300°C
Temperature at 4 mA
0 to 3000°C
0°C
Temperature at 20 mA
0 to 3000°C
1000°C
Absolute humidity of the atmosphere
0 to 689.47 kPa abs 101.33 kPa abs
8.6.5
Setting Purging
Purging is to remove condensed water in the calibration gas pipe by supplying a span calibration
gas for a given length of time before warm-up of the detector. This prevents cell breakage during
calibration due to condensed water in the pipe.
Open the solenoid valve for the automatic calibration span gas during purging and after the purge
time has elapsed, close the valve to start warm-up.
Purging is enabled when the cell temperature is 100°C or below upon power up and the purge
time is set in the range of 1 to 60 minutes.
Tag:
30
°C
4mA -Output1
4mA -Output2
Hold
Purge
F8.19.ai
Figure 8.21
Display during Purging
Содержание EXAxt ZR22G
Страница 15: ...Blank Page...
Страница 51: ...Blank Page...
Страница 77: ...Blank Page...
Страница 137: ...Blank Page...
Страница 171: ...Blank Page...
Страница 195: ...Blank Page...
Страница 197: ......
Страница 199: ......
Страница 201: ......
Страница 203: ......
Страница 205: ......
Страница 207: ......
Страница 211: ...Blank Page...