<2. Specifications>
2-13
IM 11M12A01-02E 16th Edition : Mar. 19, 2020-00
2.4
ZR402G Separate type Converter
l
Standard Specification
The ZR402G Separate type Converter can be controlled by LCD touchscreen on the converter.
Display:
LCD display of size 320 by 240 dot with touchscreen.
Output Signal:
4 to 20 mA DC, two points (maximum load resistance 550 Ω)
Contact Output Signal: Four points (one is fail-safe, normally open)
Contact Input:
Two points
Analog Input:
One point (thermal input 4-20 mA)
Automatic Calibration Output: Two points (for dedicated automatic calibration unit)
Ambient Temperature: -20 to 55°C
Storage Temperature: -30 to 70°C
Ambient Humidity:
0 to 95%RH (Non-condensing)
Power Supply Voltage: Ratings; 100 to 240 V AC, Acceptable range; 85 to 264 V AC
Power Supply Frequency: Ratings; 50/60 Hz, Acceptable range; 45 to 66 Hz
Power Consumption: Max. 300 W, approx. 100 W for ordinary use.
Maximum Distance between Detector and Converter: Conductor two-way resistance must be
10 Ω or less (when a 1.25 mm
2
cable or equivalent is used, 300 m or
less.)
Construction:
Equivalent to NEMA 4X/IP66 (with conduit holes
completely sealed with a cable gland)
Wiring Connection:
G1/2, Pg13.5, M20 x 1.5 mm, 1/2 NPT (with plug), eight holes
Installation:
Panel, wall or pipe mounting
Case:
Aluminum alloy
Paint Color:
Door: Sliver gray (Munsell 3.2PB7.4/1.2)
Case: Sliver gray (Munsell 3.2PB7.4/1.2)
Finish:
Polyurethane corrosion-resistance coating
Weight:
Approx. 6 kg
l
Functions
Display Functions: (inclused Humidity Analyzer)
Value Display; Displays values of the measured oxygen concentration, moisture
quantity, mixture ratio, etc
Graph Display; Displays trends of measured oxygen concentration, moisture quantity,
mixture ratio, etc.
Data Display; Displays various useful data for maintenance, such as cell temperature,
reference junction temperature, maximum/minimum oxygen
concentration, maximum/minimum moisture quantity or the like
Status Message; Indicates an alarm or error occurrence by flashing of the corresponding
icon. Indicates status such as warming-up, calibrating,or the like by
icons.
Alarm, Error Display; Displays alarms such as “Abnormal oxygen concentration”,
“Abnormal moisture quantity” or errors such as “Abnormal cell
e.m.f.” when any such status occurs.
Calibration Functions:
Automatic calibration; Requires the ZR40H Automatic Calibration Unit. It calibrates
automatically at specified intervals.
Содержание EXAxt ZR22G
Страница 15: ...Blank Page...
Страница 51: ...Blank Page...
Страница 77: ...Blank Page...
Страница 137: ...Blank Page...
Страница 171: ...Blank Page...
Страница 195: ...Blank Page...
Страница 197: ......
Страница 199: ......
Страница 201: ......
Страница 203: ......
Страница 205: ......
Страница 207: ......
Страница 211: ...Blank Page...