44
IM CW500-02EN
Active power P [W]
Range
Current
Voltage
96060
96061
96062
2.000 A
50.00 A
5.000 A
100.0 A
10.00 A
1000V
2000
50.00 k
5000
100.0 k
10.00 k
600.0V
30.00 k
30.00 k
3000
60.0 k
6000
Current
Voltage
96063
96064
96065
200.0 A
20.00 A
500 A
50.00 A
1000 A
100.0 A
1000V
200.0 k
20.00 k
500 k
50.00 k
1000 k
100.0 k
600.0V
120.0 k
12.00 k
300.0 k
30.00 k
600.0 k
60.00 k
Current
Voltage
96066
3000 A
1000 A
300.0 A
1000V
3000 k
1000 k
300.0 k
600.0V
1800 k
600.0 k
180.0 k
Display
digits
4 digits
Accuracy
±0.3% rdg ± 0.2% rng+ current clamp-on probe accuracy (power
factor 1, sine wave, 40 to 70 Hz)
*
The sum value is the total value of all used channels.
Power factor
influence
±1.0% rdg (40 Hz to 70 Hz, power factor 0.5)
Polarity
display
Consumption (inflow): + (no sign), regeneration (outflow):-
Equation
(
)
×
=
∑
−
=
1
0
1
n
i
ci
ci
c
A
V
n
P
c: measurement channel
i: sampling point No
*
n: Number of samples
*
For 50 Hz
,
calculation is performed on 10 waveforms (8192
points). For 60 Hz, calculation is performed on 12 waveforms (8192
points).
1P2W-1 to 4
P
1
,
P
2
,
P
3
,
P
4
,
P
sum
=
P
1
+P
2
+P
3
+P
4
1P3W(3P3W)
-1 to 2
P
1
,
P
2
,
P
sum1
=
P
1
+P
2
P
3
,
P
4
,
P
sum2
=
P
3
+P
4
P
sum
=
P
sum1
+P
sum2
3P3W3A
P
1
,
P
2
,
P
3
,
P
sum
=
P
1
+P
2
+P
3
*
Use the phase voltage.
3P4W
P
1
,
P
2
,
P
3
,
P
sum
=
P
1
+P
2
+P
3
Specifications