19
IM CW500-02EN
Output terminals
Input Terminals
Do not mistake input terminals for output terminals or vice versa.
Connectable signal wire sizes are as follows.
Compliant wire: Single wire φ1.2 mm (AWG16)
Stranded wire 1.25 mm
2
(AWG16) strand diameter
0.18 mm or more
Usable wire:
Single wire φ0.4 to φ1.2 mm (AWG26 to 16)
Stranded wire 0.2 to 1.25 mm
2
(AWG24 to 16)
Strand diameter 0.18 mm or more
Standard length of stripped wire: 11 mm
1.
Open the connector cover.
2.
While pressing the rectangular area above the appropriate terminal
with a flat-bladed screwdriver or the like, insert a signal wire.
3.
Pull out the screwdriver. The signal wire is fixed in place.
Signal wire
Flathead screwdriver
Input Terminals
Input terminals are used to monitor voltage output signals from
temperature sensors or the like. They are useful when you need to
simultaneously measure a signal output from another device and the
anomalies that occur on that power supply.
Number of channels: 2
Input resistance: Approx. 225.6 kΩ
Connecting the I/O Terminals