6.4.2 Informative messages
6.5
Accessories
IDS sensors
EC9339
Calibration error, measurement disabled
Cause: Sensor cap is missing, leaky, depleted, or defective
* Clean sensor and space between cap and sensor accord
-
ing to op.instructions
* Screw on sensor cap securely
* Repeat calibration
* Replace sensor cap
Message code
Message text
IC1339
Sensor has been successfully calibrated
* For calibration data, see calibration history
IC3339
Factory calibration has been activated. Make sure the sen-
sor operates correctly.
IC4339
Last valid calibration has been activated. Make sure the
sensor operates correctly.
IC5339
Invalid user calibration has been replaced by last valid user
calibration. Caution! Wrong measured values possible.
Carry out a new successful calibration to make sure the
sensor operates correctly.
IC6339
The invalid user calibration was replaced by the factory cal-
ibration. Caution! Wrong measured values possible. Check
whether the sensor operates correctly or carry out a new
successful calibration.
IC7339
This message shows the result of the last check (see
section 6.2.2)
IS1339
This message shows the information stored in the sensor
cap.
Message code
Message text
Description
Model
Order no.
Optical IDS oxygen sensor
FDO
®
925
201300
FDO
®
925-P
201306
Replacement sensor cap
SC-FDO® 925
201310
Содержание wtw MIQ/IDS
Страница 1: ...OPERATING MANUAL ba77277e01 01 2024 MIQ IDS IQ SENSOR NET MODULE WITH IDS INTERFACES Power OK...
Страница 2: ...MIQ IDS 2 ba77277e01 01 2024 Copyright 2024 Xylem Analytics Germany GmbH Printed in Germany...
Страница 60: ...Disposal IQ SENSOR NET 60 ba77277e01 01 2024...
Страница 61: ......