Dissolved oxygen
MIQ/IDS
30
ba77277d01
01/2024
Canceling the check
As long as the determination of the relative slope has not yet been started (Step
2:, page 29), you can quit the check with
<M>
or
<ESC>
.
You can cancel the ongoing determination of the relative slope as follows:
Press <OK>.
The sensor starts the check/
calibration. The display
switches to the measured
value display. The
CAL
indi
-
cator flashes instead of the
main measured value.
At the same time, the momen
-
tary relative slope flashes as
the secondary measured
value with the addition of
CHECK
.
The process ends automati
-
cally as soon as the mea
-
sured values meet the
criterion for the stability con
-
trol. With a great temperature
difference between the sen
-
sor and the environment, this
may take a while. Subse
-
quently, the main measured
value and temperature are
displayed.
Calibration done.
Calibration is finished.
Confirm with <OK>.
The display returns to the
measured value display.
Step 3:
Bring the sensor into the measuring position.
Once the check has been completed, the maintenance condition
has to be switched off manually. More detailed information on the
maintenance condition is given in the IQ S
ENSOR
N
ET
system
operating manual.
Display
Explanation
1.
Open the setting menu of the sensor (see section 6.3).
2.
In the
Test
menu, select the setting
abort
and then exit the setting table
with
Save and quit
.