Messages
MIQ/IDS
58
ba77277d01
01/2024
12 Messages
12.1 Explanation of the messages
This chapter contains a list of all the message codes and related message texts
that can occur in the log book of the IQ S
ENSOR
N
ET
system for the MIQ/IDS
module.
12.1.1 Error messages
12.1.2 Informative messages
Information on
the contents and structure of the log book and
the structure of the message code
is given in the L
OG
BOOK
chapter of the IQ S
ENSOR
N
ET
system oper
-
ating manual.
All message codes of the MIQ/IDS output module end with the num
-
ber "561".
Message code
Message text
EI1561
Operational voltage too low
* Check installation and cable lengths, Follow installation instructions
* Power supply module(s) overloaded, add power supply module(s)
* Check terminal and module connections
* Defective components, replace components
EI2561
Operational voltage too low, no operation possible
* Check installation and cable lengths, Follow installation instructions
* Power supply module(s) overloaded, add power supply module(s)
* Check terminal and module connections
* Defective components, replace components
Message code
Message text
EX1561
A non-permitted sensor has been connected.
* Disconnect non-permitted sensor