MIQ/IDS
Dissolved oxygen
29
ba77277d01
01/2024
Step 1:
The sensor is in the sample.
The sensor is selected in the measured value display.
The procedure is started from the measured value display with
<C> (see IQ S
ENSOR
N
ET
system operating manual).
During calibration, the sensor is in maintenance condition.
All linked outputs remain in their current status.
When operating the sensor in a flow through armature:
Remove the sensor from the flow armature.
Step 2:
Check:
Follow the instructions on the display.
Display
Explanation
* Pull the sensor out of the sam
-
ple
* Put the sensor in the calibrating
position
Place the sensor in the FDO
®
Check check and storage
vessel.
Select a procedure:
CHECK
CALIBRATION
Select the
CHECK
procedure.
Confirm with <OK>.
Leave the sensor in the check
and storage vessel long
enough so it can adapt to the
ambient temperature (min. 15
minutes).
Содержание wtw MIQ/IDS
Страница 1: ...OPERATING MANUAL ba77277e01 01 2024 MIQ IDS IQ SENSOR NET MODULE WITH IDS INTERFACES Power OK...
Страница 2: ...MIQ IDS 2 ba77277e01 01 2024 Copyright 2024 Xylem Analytics Germany GmbH Printed in Germany...
Страница 60: ...Disposal IQ SENSOR NET 60 ba77277e01 01 2024...
Страница 61: ......