Chlorine
MIQ/IDS
42
ba77277d01
01/2024
8
Chlorine
The MIQ/IDS connects IDS sensors and sensors with IDS adapter to the
IQ S
ENSOR
N
ET
.
Suitable sensors: see section 8.5 A
8.1
Measuring
Prerequisites
The IQ S
ENSOR
N
ET
system is in operation
(see IQ S
ENSOR
N
ET
system operating manual).
The sensor is connected to the MIQ/IDS module
8.2
Calibration
8.2.1 General information on calibration
Calibration adjusts the raw value measured directly in the test sample to an
independently determined reference value ("lab value"). To determine the refer
-
ence value, a sample is taken from the measuring solution and the relevant
concentration is measured (e.g. photometrically).
A common method is the DPD method for free chlorine. Photometers and test
sets for this can be found in the WTW catalog or on the Internet.
When does calibra
-
tion make sense?
Calibrating can be useful in the following cases:
If the measured values appear to be implausible
Routinely within the framework of the company quality assurance
CAUTION
Contact with the sample can lead to danger to the user!
Depending on the type of sample, suitable protective mea
-
sures must be taken (protective clothing, protective goggles,
etc.).
1. Install the sensor in a flow through armature and provide the sample flow.
or
Submerse the sensor in the sample.