![Xylem WTW DIQ/S 281 Скачать руководство пользователя страница 93](http://html1.mh-extra.com/html/xylem/wtw-diq-s-281/wtw-diq-s-281_operating-manual_3901770093.webp)
DIQ/S 281
Outputs
93
ba77244d01
08/2020
6.5
Deleting a link with an output
If a link between a current or relay output and a sensor is no longer required, you
can erase the link.
1
Open the
Einstellungen/Settings
menu with
<S>
.
2
Using
<
>
and
<OK>
, select and confirm the menu item,
System settings
->
Settings of outputs and links
.
The
Settings of outputs and links
display opens.
3
Press
<
>
to highlight the
&
column and confirm with
<OK>
.
4
Select a linked output with
<
>
and confirm with
<OK>
.
figure 6-14 Settings of outputs and links
:
Erase link
5
Using
<
>
, select
Erase link
and confirm with
<OK>
.
A safety query appears.
6
Using
<
>
, select
Erase link
and confirm with
<OK>
.
The link is deleted.