4.8.1 Maintenance condition of IQ sensors
The following diagram gives you an overview of when an IQ sensor is in the
maintenance condition.
1
Switch on the maintenance condition for the IQ sensor (see
section 4.8.2).
The display of the sensor in the measured value display flashes.
2
Pull the sensor out of the sample.
3
Perform the calibration in the laboratory, cleaning, maintenance or
repair (removing and replacing) of the sensor (about these topics, see
the component operating manual of the corresponding sensor).
4
Submerse the sensor in the sample again.
5
Wait until the measured value no longer changes.
6
Switch off the maintenance condition for the sensor (see section 4.8.3).
The display of the sensor in the measured value display no longer
flashes.