Outputs
DIQ/S 281
90
ba77244d01
08/2020
Negative charac-
teristic curve
The proportional output range begins below the initial value. If the proportional
range is undercut or exceeded, the selected behavior comes into force.
figure 6-10 Pulse width output
figure 6-11 Frequency output
1
Start value
Measured value with a minimum pulse width or switching frequency
2
End value
Measured value with a maximum pulse width or switching frequency
0
10
50
90
100
Proportional
band
t
On
t
Of f
50 %
50 %
10 %
10 %
90 %
90 %
Cycle
duration T
Time
Measured value 1
2
Pulse
v [%]
width
0
50
100
Measured value
Proportional
band
1
50 % of maximum
switching frequency
Maximum
switching frequency
Time
2
Switching frequency f [1/min]
Minumum
switching frequency