![Xylem WTW DIQ/S 281 Скачать руководство пользователя страница 62](http://html1.mh-extra.com/html/xylem/wtw-diq-s-281/wtw-diq-s-281_operating-manual_3901770062.webp)
Operation
DIQ/S 281
62
ba77244d01
08/2020
4.8.2 Switching on the maintenance condition
Switch on the maintenance condition manually when you want to clean, service
or repair (remove and replace) an IQ sensor.
= Maintenance condition ON
g m e
/
/
m
c
e
/
m
g
c
g
Measured value
display
The measured value
flashes.
Measured value display
The measured value
stops flashing.
Measured value
display
flashes
Clean
Switch on the
maintenace
condition
Continue / Cancel
Display mesage
Calibration
Step 1
Calibration
Step n
end
Maintenance
condition
ON manual
Sensor
cleaning
active
(automatic sequence)
Sensor
cleaning
OFF
Maintenance
condition
OFF manual
= Maintenance condition OFF
e
1
Call up the measured value display with
<M>
.
2
Open the
Display/Options
menu with
<OK>
.
3
Using
<
>
, highlight the menu item,
Switch maintenance
condition on/off
, and confirm with
<OK>
.
A window that informs you about the maintenance condition opens up.
4
Use
<OK>
to confirm
Continue
. The selected sensor is in the
maintenance condition. Linked outputs are frozen.
Содержание WTW DIQ/S 281
Страница 2: ...DIQ S 28X 2 ba77244d01 08 2020 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...
Страница 128: ...Messages DIQ S 281 128 ba77244e01 08 2020...
Страница 131: ......