DIQ/S 281
Outputs
117
ba76226e02
03/2023
Canceling the
cleaning
A running cleaning cycle is canceled:
Automatic
– if the sensor switches to the inactive condition during the cleaning cycle
Manual
– By pressing the
<C>
key
– By switching on the maintenance condition
Each time the cleaning cycle is canceled, the relay opens immediately.
If the cleaning cycle is canceled automatically, the outputs linked to the sensor
are released immediately.
If the cleaning cycle is canceled manually, the sensor is in the maintenance
condition. The linked outputs are only released after the maintenance condition
was terminated manually.
1 Reference time
Begin of a cleaning cycle (
t1
)
Begin of the specified
Cleaning duration
(
t1a
)
2
End of the specified
Cleaning duration
(
t1a
)
Begin of the specified
Adjustment time
(
t1b
)
3
End of the specified
Adjustment time
(
t1b
)
End of the cleaning cycle (
t1
)
4
Reference time ±
Cleaning interval
(
t2
)
Begin of a cleaning cycle
relay
condition
open
closed
00:00
Time
3
1
2
t1
t1
t1
t1a t1b
4
4
t2
t2
t2
t2
1
04:00
12:00
24:00
20:00