DIQ/S 281
Outputs
101
ba76226e02
03/2023
switched more often or less often.
Characteristic
curves
Through the selection of the
Start value
and
End value
, the proportional output
can be operated with a positive or negative characteristic curve.
Positive characteristic curve:
Select the
End value
to be greater than the
Start value
.
The turn-on duration or frequency increases with an increasing measured
value (see page 103).
Negative characteristic curve:
Select the
End value
to be smaller than the
Start value
.
The turn-on duration or frequency decreases with an increasing measured
value (see page 104).
The maximum values for the pulse width or switching frequency are assigned to
the
End value
value and the minimum values for turn-on duration or frequency
are assigned to the
Start value
value.
figure 6-5
Relay output of frequency output
While the selected switching duration (t
on
= 0.3 s) always remains con-
stant, the switching frequency at which the relay switches changes
depending on the measured value.
If the measured value is at the end of the proportional range (
End
value
), the switching frequency is higher.
If the measured value is at the beginning of the proportional range
(
Start value
), the switching frequency is low.
Time [s]
t
= 0.3 s
On
Relay
Off
On
T