XO FLEX – Installation instructions
Installation
32
The post for the operating light arm and the outer cylinder may be fitted
simultaneously in clinics with ceiling height lower than 255 cm. The outer cylinder
can be supported by the lower plastic guide ring on the post during assembly.
FITTING THE LIGHT ARM
The easiest way to fit the light arm is to start by fitting the arm and then routing the cables down
through the light arm post.
1.
Lower the light arm carefully onto the light arm post – be careful not to tilt the arm and squeeze
the bearings. See Figure 40.
2.
Route the cables through the light arm post and pull them from the hole of the electronics side
of the unit stand. If necessary, use a piece of hard wire as a “search spring”. See Figure 41.
Figure 40 – Fitting the light arm
Figure 41 – Cables from the light