Appendix A—Output Specifications
89
Settling Time: -------Less than 2 seconds to within 10ppm of 50pA.
Drive Capability: ---Greater than 18V compliance ability, stable into active
and/or inductive loads.
Overload:-------------Output fully protected against open circuits, or applied
voltages less than 50V.
Output Impedance: -Greater than 1011
Ω
Long-Term Effects: Add 60ppm of output to 1 year specification for three
year accuracy
Voltage Measurement Specifications
Voltage measurement capabilities are present in the 2000M instrument. To
measure voltage, access the instrument’s “Measurement” mode.
In Voltage Measurement mode (-0.15 to +0.15V):
•
A 90 day absolute accuracy is 0.01% +1µV +3ppm +1
µ
V throughout the
temperature range.
•
A 1 year absolute accuracy is 0.02% +1µV +3ppm +1
µ
V throughout the
temperature range.
Thermocouple Specifications
The specifications for Thermocouple simulation and measurements are
shown in Table 9. All temperatures are given in ºC, the desired temperature
may be defined in the instrument in either ºC or ºF units. The required output
voltage is internally computed from the entered temperatures from N.I.S.T.
look-up tables for the respective thermocouple type with interpolation.
Note:
Both cold and hot junction temperatures may be entered.
Содержание 2000I
Страница 1: ...USER GUIDE 2000I 2000M DC Power Source Instruments...
Страница 2: ......
Страница 10: ...10 2000I 2000M User Guide Rev B...
Страница 20: ...20 2000I 2000M User Guide Rev B...
Страница 44: ...44 2000I 2000M User Guide Rev B...
Страница 76: ...76 2000I 2000M User Guide Rev B...
Страница 96: ...2000I 2000M Instrument s User Guide Revision B 96...