Calibration and Maintenance
75
present prior to disassembly. Ensure the nylon washers are included
when reattaching the Analog board.
Warning:
The printed circuit boards are continuously powered by the
battery, use extreme care in handling them. Reconnect the battery cables after
reattachment of the boards.
5.
Following complete reassembly, fully charge the replaced battery, in
accordance with the charging procedures in the
Battery Charging
chapter.
6.
Perform an internal and external recalibration on instrument after
charging in accordance with the calibration procedures.