fits or export a zip file containing the traces and the analysis log file (see
). This plot
can be zoomed/panned, as described in
c)
Navigation Buttons
Navigation buttons: There are two pairs of navigation buttons. The left and right exterior buttons go
to the first and last events, respectively. The home/end and up/down arrows on the keyboard have the
same effect, as does pressing the Ctrl+F or Ctrl+L keys. The interior buttons scroll through events
sequentially, in the order they were captured, one at a time. The left/right arrows on the keyboard have
the same effect, as does pressing the Ctrl+N or Ctrl+P keys. The trigger number of the current event is
displayed beneath the navigation buttons, and you can enter a number into it in order to display that
event.
d)
Event Classifications and Counts
Displays the classifications and number of each event type observed in the measurement. Clicking
on an event class will toggle the visibility of the associated events in the Risetime vs. Anode Energy
scatterplot and determine whether it will be skipped while scrolling. This is a convenient way to closely
inspect the traces in a certain event class without having to scroll through all the rest. All event classes
are initially toggled on. The event classes with a grey background aren’t displayed in the Risetime vs.
Anode Energy scatterplot plot because they don’t have a well-defined risetime, but they will be scrolled
through. The numbers and rates of events can be a useful diagnostic tool, for more information see
e)
Log button
Shows the Alpha Analysis Engine (AAE) log file instead of the trace. Used for diagnostics. Underneath
is the trigger time for the displayed event.
f)
Open button
Opens a file for re-analysis. Click the button, navigate to where the file is stored, and open the file.
The filename will now be populated in the text box next to the button. This button, along with the text
box and the other buttons in this row, will disappear during data acquisition.
g)
Analyze button
After opening a file, click to analyze it. CounterMeasure will analyze the file as though the data were
being actively taken, and display the results as it analyzes. This button will be disabled when there is no
file selected and disappear during data acquisition.
h)
Parameters Button
Opens the
. Like the other buttons in this row, it will disappear during data
acquisition.