UltraLo-1800 Alpha Particle Counter
XIA LLC
Page 27
Rear View
1.
Air inlet: Purges the outer chamber with air in case of any argon leaks.
2.
Power connector: brings power to the electronics.
3.
USB cable opening.
4.
Gas inlet: Allows argon into the Active Volume.
5.
Pressure gauge: Allows reading the internal pressure to ensure that gas exit and the door
seal are working as intended.
6.
Gas outlet: Provides a controlled exit of argon from the Active Volume.
7.
Check valve: Provides overpressure exhaust relief.
8.
Motor enclosure: Powers the tray mechanism.
Front View
9.
Door: Forms seal to keep argon pure, allows samples to be introduced.
10.
Side fairings: Aesthetic coverings.
11.
Vibration isolating feet: Damp mechanical vibrations.
Figure IV-2: Left: rear view of counting module. Right: front view of counting module.