User Manual
Wildcat
Company confidential.
This document is the property of Xenics. It may not be reproduced
–
completely or partially
–
or passed to a third
party without written permission from Xenics.
Xenics nv
T +32 16 38 99 00
Doc Ref: ENG-2020-UMN001
Ambachtenlaan 44
F +32 16 38 99 01
Issue: R001
BE-3001 Leuven
Belgium
www.xenics.com
Date:
XF-104_03/20-12-2012
Page 47 of 58
Storing a custom calibration in camera memory
Clearing or overwriting existing calibration sets stored in camera memory is
irreversible.
There are 10 set slots reserved in camera memory for storing NUC corrections, labeled
NUC set 0..9
. This functionality may be found in the
Settings
– Storage
tab in Xeneth,
under the
Calibration set control
section.
Selector
allows you to select the calibration memory slot to modify.
When a slot is selected,
Slot info
indicates whether a calibration set already occupies that
slot.
If a slot is already occupied and needs to be overwritten, it first needs to be cleared by
selecting
Clear
and pressing the
Execute
button
.
To upload a calibration set into the selected memory slot:
1. Press the green upload arrow
next to
Store
.
2. Select the corresponding calibration file in the file dialog.
3. Wait until the transfer progress bar closes and the file has transferred to the camera.
The corresponding camera registers for on-board calibrations may be found in Chapter
9.1.9.
!