Calibration
Setup and Equipment
182
Operating Manual for XDC Series Power Supply
Setup and Equipment
•
6 digit DVM
•
current shunt
•
variable load
•
0-10 V DC power supply (analog programming interface)
•
Load wiring sized for the maximum available output current. See Table 2.4, on
page 48.
To set up to calibrate output voltage and current:
Connect a load to the output of the power supply and a current shunt in series.
You will need to use the DVM to measure both the voltage at the output of the power
supply and the voltage across the shunt. You will need to convert the voltage across
the shunt to a current measurement.
To set up to calibrate the analog programming interface, you will need a 0-10V
power supply to provide a programming signal to the voltage and current
programming lines. You will also need to connect the DVM to measure the signal at
the readback lines.
Содержание XDC
Страница 2: ......
Страница 3: ...Operating Manual for XDC 6000 Watt and 12000 Watt Series Digital Programmable DC Power Supply ...
Страница 22: ...List of Figures xx Operating Manual for XDC Series Power Supply ...
Страница 34: ...About The XDC Power Supply Overview of Operation 32 Operating Manual for XDC Series Power Supply ...
Страница 57: ...Installation Load Connections Release 3 0 55 Figure 2 11Output for 12000 Watt units High Voltage 300 600V ...
Страница 174: ...Remote Operation SCPI Commands for Digital Interfaces 172 Operating Manual for XDC Series Power Supply ...
Страница 180: ...Current Sharing 6000 Watt only Operation 178 Operating Manual for XDC Series Power Supply ...
Страница 216: ...SCPI Command Reference Expressions 214 Operating Manual for XDC Series Power Supply ...
Страница 228: ...GPIB Performance Specifications 226 Operating Manual for XDC Series Power Supply ...
Страница 253: ......