I N S T R U M E N T O P E R A T I O N
5-27
Compare
The Compare function is a quick method for comparing
measurements without storing the data. After entering the
mode, the first measurement is set as the standard and each
measurement thereafter is compared to it. The standard can be
remeasured whenever desired. Measured data can be
automatically transmitted from the RS-232 port if Auto Xmt is
activated in the Instrument Configuration.
To access the Compare function:
1.
Use the Tab keys
$@
to highlight
Compare
.
–
M A I N M E N U
–
Q A
A n a l y z e
C o m p a r e
S t r e n g t h
O p a c i t y
↓
X — R i t e
9 3 9
— — — — — —
X X X X
* * * * * *
2.
Press the Enter key
#
to access the Compare function.
S t a n d a r d
C o m p a r e
<
M e a s u r e S t a n d a r d
>
L*a*b*
D65/10
......
....
L* 0.00
a* 0.00
b* 0.00
3.
Measure the standard to be used for comparison.
S a m p l e
C o m p a r e
<
M e a s u r e S a m p l e
>
L*a*b*
D65/10
......
....
∆
L* +0.00
L* 81.17
∆
a* —0.00
a* +6.39
∆
b* +0.00
b*+83.05
∆
E* 0.00
4.
Standard measurement data is displayed. Measure the sample
or press the Enter key
#
with
Sample
highlighted to toggle
to
Standard
. Another standard can then be measured.
S a m p l e
C o m p a r e
<
M e a s u r e S a m p l e
>
L*a*b*
D65/10
......
....
∆
L* –0.08
L* 81.09
∆
a* —0.00
a* +6.39
∆
b* –1.08
b*+81.97
∆
E* 1.08
Measurement instruction
Difference Data
Sample Data
Difference data appears
as all zeros before
measuring a sample
Measurement instruction
Standard Data
Содержание 939
Страница 1: ...939 S P E C T R O D E N S I T O M E T E R Operator s Manual ...
Страница 25: ...C H A P T E R T W O 2 8 ...
Страница 29: ...C H A P T E R T H R E E 3 4 ...
Страница 57: ...C H A P T E R F O U R 4 28 ...
Страница 101: ...C H A P T E R S I X 6 4 ...
Страница 107: ...C H A P T E R S E V E N 7 6 ...