Error Messages
Errors encountered during a measurement are displayed in
the User Dialog. All errors are accompanied by a long
beep. Any errors (except
<
Low Battery
>
) encountered
during a measurement cancel that measurement; and the
data displayed is from the prior measurement.
User Dialog Errors:
<
Measure Aborted
>
Displays with an incomplete measurement.
Instrument was release too soon.
<
Invalid Measure
>
Measure did not complete successfully. Try
again. Could be a hardware failure if it
occurs again.
<
Needs Cal!
>
Displays when calibration is required.
<
Cal Failed!
>
Calibration failed. Make sure the instrument
is properly positioned on the reference.
<
Cal Aborted!
>
Displays with an incomplete cal
measurement. Instrument was release too
soon.
<
Low Battery
>
This warning appears when the battery falls
below approximately 25% of full charge.
Measurements are still possible, but the
battery should be charged soon.
<
CHARGE BATTERY
>
Displays when not enough battery power
remains to take measurement. The current
measurement is aborted.
<
CHECK BATTERY
>
Battery not installed, disconnected, or dead.
Unit will not allow any measurements.
<
50% Dot Fail
>
Did not measure a true 50% dot tint in 50%
Dot Cal procedure.
<
WRONG CHARGER
>
Wrong charger connected to unit.
MP Failure
This error "pops up" instead of appearing in
the User Dialog. It is generated when trying
to change the configuration, but the unit
cannot make the change permanent. Try
again.
Содержание 500 Series
Страница 1: ...500 Series S P E C T R O D E N S I T O M E T E R Operator s Manual ...
Страница 2: ......
Страница 20: ...C H A P T E R O N E 1 8 ...
Страница 54: ...C H A P T E R F O U R 4 24 ...
Страница 112: ...C H A P T E R F I V E 5 58 ...
Страница 120: ...C H A P T E R S I X 6 8 ...
Страница 123: ......