C H A P T E R F I V E
5
-
56
Electronic Function Selection (518, 528, 530 only)
EFS automatically recognizes the patch type measured
without manually selecting the function. The "smart"
recognition of Paper, Dot, Solid, PC, Density, Trap, H/G,
and Color is contained in this function. Color only
appears on the 528 and 530 instruments.
NOTE:
If the highlight in the measurement list does not move to the
desired function after a measurement, use the Tab keys
$@
to move
the highlight to the function before releasing the instrument.
Setting Options
Pressing the Enter key
#
with the Options menu highlighted
opens the EFS Options menu. The OP Setting, Ref 3 Setting,
and Mode options are selected under this menu item.
OP Setting
The Overprint option allows you to choose Trap, H/G, or
Color. When the option is set to Trap, the instrument
automatically switches to the Trap function when an
overprint is measured. When the option is set to H/G, the
highlight moves to Hue/Grayness when an overprint is
measured and displays H/G data. And when the option is set
to Color, the highlight moves to Color when an overprint is
measured and displays color data.
Ref 3 Setting
The Ref 3 Setting option allows you to select between Dot
and Print Contrast (PC). When set to PC, Print Contrast
data appears after a 75% tint measurement. And when set
to Dot, Dot data appears after a tint measurement.
Mode
The Mode allows you to select between Absolute and
–Paper.
E F S O p t i o n s :
O P S e t t i n g : T r a p
R e f 3 S e t t i n g : D o t
M o d e : A b s o l u t e
<
E d i t O p t i o n s
>
Trap, H/G, Color
Dot or PC
Absolute or -Paper
Содержание 500 Series
Страница 1: ...500 Series S P E C T R O D E N S I T O M E T E R Operator s Manual ...
Страница 2: ......
Страница 20: ...C H A P T E R O N E 1 8 ...
Страница 54: ...C H A P T E R F O U R 4 24 ...
Страница 112: ...C H A P T E R F I V E 5 58 ...
Страница 120: ...C H A P T E R S I X 6 8 ...
Страница 123: ......