S E R V I C E A N D G E N E R A L M A I N T E N A N C E
6-5
5.
Press the tool down (approx. 12mm) on the aperture
with slight pressure. A low, click is heard when
properly seated.
6.
Carefully lift aperture tool upward. The old aperture
will be attached to the tool fingers. Remove old
aperture from tool and set aside.
7.
Position new aperture over opening with tabs aligned
on the "left" and "right" sides of the instrument.
8.
Slide aperture into opening. Finish pressing aperture
into position using the other end of the extraction
tool. A click is heard when the aperture is properly
seated.
Tabs
Aperture