Measuring/Editing a Color Reference
The reference function is used to enter color reference
data into the instrument using a sequence or match
method. Up to 16 references can be stored and accessed in
the instrument. Color reference values are then compared
to color measurements and the difference displayed.
To measure a reference:
1.
If not selected, press the Tab Up key
$
or Tab Down
key
@
to highlight
Reference
and press Enter key
#
.
NOTE:
Reference
will not appear in the measurement list unless
COLOR–REF01
is selected as the active color mode. Refer to Color
Measurement Mode earlier in this section.
2.
Press the Tab Up key
$
to move highlight to the
reference measurement method (
Seq
or
Match
).
•
The
Sequence
(Seq) method automatically
increments the reference location as you measure.
For example, when the highlight is on
Ref 01
, and a
measurement is taken, the data is set for reference
one and the highlight advances to
Ref 02
. This
occurs for all reference measurements taken.
•
The
Match
method is used to update an existing
reference. The measurement automatically replaces
data in the location that has the closest match (lowest
Δ
E) to the measurement.
C O L O R
-
R E F 0 1
O p t i o n s
S a m p l e
L* 31.06
a* —0.05
b*—33.12
<
M o d i f y R e f
>
D 5 0 / 2
R e f e r e n c e s
R e f e r e n c e s
S e q
R e f 0 1
R e f 0 2
R e f 0 3
R e f 0 4
^
L* 89.45
a* 1.46
b* 21.53
<
M e a s u r e R e f
>
D50/2
Reference
measurement method
Содержание 500 Series
Страница 1: ...500 Series S P E C T R O D E N S I T O M E T E R Operator s Manual ...
Страница 2: ......
Страница 20: ...C H A P T E R O N E 1 8 ...
Страница 54: ...C H A P T E R F O U R 4 24 ...
Страница 112: ...C H A P T E R F I V E 5 58 ...
Страница 120: ...C H A P T E R S I X 6 8 ...
Страница 123: ......