C H A P T E R F I V E
5
-
20
Measuring References
The reference function is used to measure reference data into a
selected group. A total of 1,424 references can be stored in the
instrument. The default reference name that is generated after
a measurement (i.e., Ref0001) can be change to a specific
color name or code.
NOTE:
Reference databases that are downloaded using ColorMail
Express cannot be edited or deleted. The word <LOCKED> appears
is the user dialog, indicating the reference cannot be changed.
M A T C H : G r o u p 1
O p t i o n s
S a m p l e
M a t c h
L* 34.94
a* —1.98
b* —25.65
<
V i e w Re f s
>
D 5 0 / 2
To create a reference:
1.
Make sure that the active group you want to store the
reference in is displayed (see Setting Options for
selection procedure).
2.
Press the Tab Up key
$
or Tab Down key
@
to
highlight
References
and press the Enter key
#
.
R E F : G r o u p 1
O p t i o n s
A d d N e w
<
M e a s u re R ef
>
D 5 0 / 2
3.
Make sure
Add New
is highlighted. Center the target
window over the reference and lower the instrument to
take a reading.
R E F : G r o u p 1 O p t i o n s
A d d N e w
R e f 0 0 0 1
Ref0001
L* 33.44
a*
–
1.27
b*
–
29.92
Delete
<
V i e w Re f s
>
D 5 0 / 2
The instrument automatically numbers (Ref0001)
each reference by default.
R e f e r e n c e s
Содержание 500 Series
Страница 1: ...500 Series S P E C T R O D E N S I T O M E T E R Operator s Manual ...
Страница 2: ......
Страница 20: ...C H A P T E R O N E 1 8 ...
Страница 54: ...C H A P T E R F O U R 4 24 ...
Страница 112: ...C H A P T E R F I V E 5 58 ...
Страница 120: ...C H A P T E R S I X 6 8 ...
Страница 123: ......