What to do if ...
44
Cause
Remedy
– Electrolyte depleted
– Replace the membrane
head and calibrate
according to the
operating manual
– Counter electrode
contaminated
– Clean the electrodes
Cause
Remedy
– The membrane head is
loose
– Screw the membrane
head tight.
– There is a hole in the
membrane
– Replace the membrane
head and recalibrate
according to the
operating manual, if
necessary clean the
electrodes
Cause
Remedy
– Defective temperature
sensor
– Send probe to WTW
The Sens Reg
message appears
The leakage
display appears
Wrong temperature
display