Maintenance, cleaning, disposal, spare parts
30
Caution
Never immerse the reference electrode into the cleaning
solution! The reference electrode might be destroyed and
the probe defective. If the RA 600 is used correctly,
however, there is no danger for the probe.
Allow the cleaning solution to react for max. 1 hour.
6 Screw the safety cap with the probe on the screw
cap.
7 Screw off the protection cap.
8 Thoroughly rinse the electrode - with the safety cap
put on - with deionized water.