Technical data
49
7.2
Measuring conditions
0 - 60 mg/L O
2
0 - 600 % saturation (
≈
0 - 1200 mbar pO
2
)
Measuring range
for initial commissioning or after replacing the
electrolyte: at least 60 min;
after short polarization breaks: 15 to 60 min depending
on the duration of the break
Operation:
0 °C ... + 50 °C
Storage:
- 5 °C ... + 50 °C
10 bar
min. 10 cm; max. 100 m depth
7-pin plug
Connection
Probe - cable:
IP 68 (100 m)
Cable - plug:
IP 65
Plug - instrument: IP 65
any, if the minimum flow is granted
≥
0.5 cm/s for 5 % measuring accuracy
≥
23 cm/s for < 1 % measuring accuracy
Polarization time
Temperature range
Max. admissible
overpressure
Immersion depth
Connection
Operating position
Flow to the probe