Technical data
51
7.4
Electrical data
Top view of the plug
Warning
All voltages must be protective low voltages or safety
extra low voltages without shock-hazard according to -
IEC 1010 or UL3101.
All currents to the probe must not exceed 8 A even in case
of a short-circuit.
Pin Assignment
Voltage
maximum
current
1
2
3
4
5
6
7
Ub+
0V
NTC (30kOhm)
NTC (30kOhm)
Ub-
Reg/Leak
(output signal)
pO
2
(output signal)
+ 6.5 V ... + 8.5 V
---
< 3.5 V
< 3.5 V
- 6.5 V ... - 8.5 V
- 8.5 V ... + 8.5 V
0 V ... + 8.5 V
< 1mA
---
< 0.15mA
< 0.15mA
< 1 mA
---
---
Pin assignment