Maintenance, cleaning, disposal, spare parts
25
4.2
Cleaning the electrodes
Note
Counter electrode and reference electrodes always show a
coloring. This is necessary for the operation of the probe. It
is not a contamination.
Cleaning is only required if the slope is too high or too low
(the probe cannot be calibrated), and if this cannot be
remedied by replacing the membrane head and the
electrolyte solution.
1 Disconnect the probe from the instrument.
2 Screw off the membrane head.