What to do if...
OxiTop
®
Control OC 100
74
ba31114e06
06/2012
Which measuring head
belongs to which
measurement curve?
Prepare the relation separate curve – number of
measuring head; see section 5.5 E
Read number of measuring head out of headline
Have measuring head shown; see section 5.4.1 S
The precision and
sealing of the system
sample bottles plus
measuring head
should be checked
Measuring head check
(single check)
(To which running
sample does the
selected measuring
head belong?)
Single check method:
Menu
GLP/TOOLS
- Check
-
info
The controller guides you further through the operation.
All single information on the state of the measuring heads is listed, i.e. the
measuring head responds.
The check can be undertaken at any time without affecting the measure-
ment that is running.
If no reaction at all can be determined, the batteries should be changed and
the measuring head check repeated.
The measuring precision cannot be checked with this!
The measuring
precision of the
measuring head
should be checked.
Perform measuring
head reaction test
Menu
GLP/TOOLS
- Check
- All
:
All optically attainable measuring heads must flash for approx. 5 seconds
independent of their status.
The check can be undertaken at any time without affecting the measure-
ment that is running.
This test only checks the reaction to commands.
If no measuring head reacts, the controller should be checked.
Cause
Remedy
Содержание OxiTop C
Страница 6: ...Contents OxiTop Control OC 100 6 ba31114e06 06 2012...
Страница 14: ...Commissioning OxiTop Control OC 100 14 ba31114e06 06 2012...
Страница 55: ...OxiTop Control OC 100 GLP TOOLS ba31114e06 06 2012 55...
Страница 60: ...Transmitting data OxiTop Control OC 100 60 ba31114e06 06 2012...
Страница 80: ...Index OxiTop Control OC 100 80 ba31114e06 06 2012...
Страница 81: ......